ASTM D6978 Chemotherapy Glove Testing: How to Read Breakthrough-Time Data

The label “nitrile” does not tell you how long a glove will resist a chemotherapy drug. Two nitrile gloves can differ in formulation, thickness and film uniformity, and one drug can move through the same glove much faster than another.

ASTM D6978 provides a controlled way to measure that behavior. It exposes medical-glove material to specified chemotherapy drugs and records the minimum breakthrough detection time for each one. The result is drug-specific data, not a single universal “chemo-proof” rating.

Laboratory scientist testing purple nitrile glove material in a chemical permeation cell

What ASTM D6978 covers

ASTM D6978-05(2023) is the current standard practice for assessing the resistance of medical-glove materials to permeation by potentially hazardous chemotherapy drugs under continuous-contact conditions. The FDA recognizes this edition, and its medical-device classification for chemotherapy-labeled gloves describes testing such as ASTM D6978 as evidence for permeation and breakthrough resistance.

The word permeation has a precise meaning. Drug molecules are absorbed into the glove material, diffuse through it at a molecular level and emerge on the other side. The glove may still look completely intact.

Failure modeWhat happensCan the glove still look intact?
PenetrationLiquid passes through a hole, seam or visible defectSometimes
PermeationMolecules move through the polymer itselfYes
DegradationThe chemical changes the glove, causing swelling, softening, hardening or loss of strengthOften, but not always

ASTM D6978 focuses on permeation. It does not replace inspection for holes, mechanical testing or evaluation of degradation.

How the permeation cell works

A specimen cut from the glove separates two chambers in a permeation cell. The outside surface faces the chemotherapy-drug solution. The collection side contains a medium that captures any drug reaching the inner surface.

Samples from the collection medium are analyzed over time using a method sensitive to the drug being tested. The laboratory calculates when the measured permeation rate reaches the standard’s reporting threshold.

The test conditions are deliberately demanding:

ParameterASTM D6978 approachWhy it matters
ContactContinuousRepresents sustained exposure rather than a brief splash
Temperature35 +/- 2 degrees CApproximates the warmer conditions of a glove worn close to skin
Specimen locationThinnest area from the palm or cuffAvoids presenting a stronger result from a conveniently thick area
Standardized breakthrough rate0.01 microgram/cm2/min (10 ng/cm2/min)Uses a sensitive threshold because chemotherapy drugs can be hazardous at low exposure levels
Maximum test duration240 minutesResults beyond this are reported as greater than 240 minutes, not as infinite resistance

The specimen-location rule matters because disposable gloves do not have perfectly uniform thickness. Fingertips, palms and cuffs can differ. A result from the thinnest relevant area gives a more conservative representation of the finished glove.

What “breakthrough time” actually means

Breakthrough time is not the moment a visible drop appears. It is the elapsed time until the drug’s permeation rate reaches a defined analytical threshold.

Under ASTM D6978, the standardized breakthrough detection time is associated with a permeation rate of 10 nanograms per square centimetre per minute. The number is therefore a laboratory endpoint tied to a known area and rate.

Consider this simplified example:

Test timeMeasured permeation rateInterpretation
30 minBelow detectionNo reported breakthrough yet
60 min4 ng/cm2/minBelow the 10 ng/cm2/min threshold
90 min12 ng/cm2/minStandardized breakthrough has occurred by this interval

This example is illustrative, not an ArmorHive product result. Actual reports should state the sampling schedule and how the minimum breakthrough time was determined from replicate specimens.

A result of “>240 minutes” means the threshold was not reached during the four-hour test. It does not mean the glove can be worn indefinitely or that no molecule crossed the film.

Which chemotherapy drugs are included

ASTM D6978 establishes seven required challenge drugs and allows two additional drugs selected by the manufacturer. Drug identity and concentration both matter because permeation behavior changes with the test solution.

Required drugTest concentration commonly specified
Carmustine3.3 mg/mL
Cyclophosphamide20 mg/mL
Doxorubicin hydrochloride2 mg/mL
Etoposide20 mg/mL
Fluorouracil50 mg/mL
Paclitaxel6 mg/mL
Thiotepa10 mg/mL
Two additional drugsSelected and concentration documented by the manufacturer

A glove can exceed 240 minutes for several drugs yet show a much shorter time for carmustine or thiotepa. FDA 510(k) summaries provide real examples of this pattern. That does not make one drug result an error; it demonstrates why the report must be read drug by drug.

Why carmustine and thiotepa often stand out

Permeation depends on the interaction between the drug formulation and the glove polymer. Molecular size, polarity, solvent system, concentration and affinity for the polymer all influence absorption and diffusion.

Carmustine and thiotepa frequently produce shorter breakthrough times in published glove data. A report that lists only the best-performing drugs can therefore create a misleading impression.

The practical rule is:

> Never transfer a breakthrough time from one drug to another, even when both are tested on the same glove.

The same rule applies to concentration. A test performed at one documented concentration does not automatically establish performance for a different formulation or mixture.

ASTM D6978 versus ASTM F739

ASTM F739 is a general method for measuring permeation of liquids and gases through protective-clothing materials. ASTM D6978 is tailored to medical gloves and chemotherapy drugs.

Comparison pointASTM D6978ASTM F739
Primary useChemotherapy-drug permeation through medical glovesGeneral chemical permeation through protective materials
Test temperature35 +/- 2 degrees CConditions selected under the method and report
Breakthrough-rate threshold10 ng/cm2/minCommon standardized threshold of 100 ng/cm2/min under current general reporting practice
SpecimenThinnest palm or cuff areaRepresentative protective-material specimen
Reporting emphasisMinimum breakthrough time for each drugBreakthrough and permeation parameters for the chemical/material pair

The D6978 threshold is ten times lower than 100 ng/cm2/min. This more sensitive endpoint reflects the hazardous nature of antineoplastic drugs. A general chemical-resistance result should not be substituted for drug-specific D6978 data when a chemotherapy claim is required.

How to read a test report correctly

A useful ASTM D6978 report should identify:

  • the exact glove model, material, size, color and lot;
  • measured thickness at the tested specimen location;
  • the standard edition;
  • every drug name and its test concentration;
  • the number of specimens or replicates;
  • individual or minimum breakthrough times;
  • results reported as minutes or greater than 240 minutes;
  • the analytical method and detection capability; and
  • the laboratory and test date.

Minimum values matter more than a convenient average. If three specimens break through at 38, 52 and 61 minutes, the test report should not allow a reader to mistake the 50-minute average for the earliest observed protection limit.

The report should also connect the tested samples to production. A test on a thicker development sample does not automatically support a thinner commercial glove, even if both are described as nitrile.

Test time is not the same as safe wear time

Laboratory breakthrough time is one input to a risk assessment, not a ready-made glove-change schedule. Real use adds factors that a flat, stationary permeation cell does not reproduce:

Workplace factorPossible effect
Stretch over fingers and knucklesLocally reduces film thickness
Flexing and grippingApplies repeated mechanical stress
Abrasion or small puncturesCreates new pathways through the film
Contact with alcohol or cleaning productsMay alter the polymer before drug exposure
Drug mixtures or excipientsCan behave differently from the tested solution
Heat and perspirationChange the conditions at the hand-glove interface

Policies for handling hazardous drugs may also require double gloving and scheduled changes. Those decisions should follow the healthcare facility’s risk assessment, product instructions and applicable regulations or professional guidance, not a blog or a single breakthrough number.

What “tested for use with chemotherapy drugs” should mean

The phrase should be supported by product-specific ASTM D6978 results. It should not mean merely that the glove is nitrile, thick, powder-free or resistant to an unrelated solvent.

A transparent claim names the drugs tested and provides their minimum breakthrough times. It also highlights short results rather than hiding them behind a broad statement.

The central lesson is straightforward: ASTM D6978 measures drug-specific molecular permeation under controlled continuous contact. The earliest result for the relevant drug matters, and a value greater than 240 minutes is only a statement about the four-hour test window. Proper glove selection combines that evidence with task conditions, inspection, changing practices and the complete hazardous-drug program.