The label “nitrile” does not tell you how long a glove will resist a chemotherapy drug. Two nitrile gloves can differ in formulation, thickness and film uniformity, and one drug can move through the same glove much faster than another.
ASTM D6978 provides a controlled way to measure that behavior. It exposes medical-glove material to specified chemotherapy drugs and records the minimum breakthrough detection time for each one. The result is drug-specific data, not a single universal “chemo-proof” rating.

What ASTM D6978 covers
ASTM D6978-05(2023) is the current standard practice for assessing the resistance of medical-glove materials to permeation by potentially hazardous chemotherapy drugs under continuous-contact conditions. The FDA recognizes this edition, and its medical-device classification for chemotherapy-labeled gloves describes testing such as ASTM D6978 as evidence for permeation and breakthrough resistance.
The word permeation has a precise meaning. Drug molecules are absorbed into the glove material, diffuse through it at a molecular level and emerge on the other side. The glove may still look completely intact.
| Failure mode | What happens | Can the glove still look intact? |
|---|---|---|
| Penetration | Liquid passes through a hole, seam or visible defect | Sometimes |
| Permeation | Molecules move through the polymer itself | Yes |
| Degradation | The chemical changes the glove, causing swelling, softening, hardening or loss of strength | Often, but not always |
ASTM D6978 focuses on permeation. It does not replace inspection for holes, mechanical testing or evaluation of degradation.
How the permeation cell works
A specimen cut from the glove separates two chambers in a permeation cell. The outside surface faces the chemotherapy-drug solution. The collection side contains a medium that captures any drug reaching the inner surface.
Samples from the collection medium are analyzed over time using a method sensitive to the drug being tested. The laboratory calculates when the measured permeation rate reaches the standard’s reporting threshold.
The test conditions are deliberately demanding:
| Parameter | ASTM D6978 approach | Why it matters |
|---|---|---|
| Contact | Continuous | Represents sustained exposure rather than a brief splash |
| Temperature | 35 +/- 2 degrees C | Approximates the warmer conditions of a glove worn close to skin |
| Specimen location | Thinnest area from the palm or cuff | Avoids presenting a stronger result from a conveniently thick area |
| Standardized breakthrough rate | 0.01 microgram/cm2/min (10 ng/cm2/min) | Uses a sensitive threshold because chemotherapy drugs can be hazardous at low exposure levels |
| Maximum test duration | 240 minutes | Results beyond this are reported as greater than 240 minutes, not as infinite resistance |
The specimen-location rule matters because disposable gloves do not have perfectly uniform thickness. Fingertips, palms and cuffs can differ. A result from the thinnest relevant area gives a more conservative representation of the finished glove.
What “breakthrough time” actually means
Breakthrough time is not the moment a visible drop appears. It is the elapsed time until the drug’s permeation rate reaches a defined analytical threshold.
Under ASTM D6978, the standardized breakthrough detection time is associated with a permeation rate of 10 nanograms per square centimetre per minute. The number is therefore a laboratory endpoint tied to a known area and rate.
Consider this simplified example:
| Test time | Measured permeation rate | Interpretation |
|---|---|---|
| 30 min | Below detection | No reported breakthrough yet |
| 60 min | 4 ng/cm2/min | Below the 10 ng/cm2/min threshold |
| 90 min | 12 ng/cm2/min | Standardized breakthrough has occurred by this interval |
This example is illustrative, not an ArmorHive product result. Actual reports should state the sampling schedule and how the minimum breakthrough time was determined from replicate specimens.
A result of “>240 minutes” means the threshold was not reached during the four-hour test. It does not mean the glove can be worn indefinitely or that no molecule crossed the film.
Which chemotherapy drugs are included
ASTM D6978 establishes seven required challenge drugs and allows two additional drugs selected by the manufacturer. Drug identity and concentration both matter because permeation behavior changes with the test solution.
| Required drug | Test concentration commonly specified |
|---|---|
| Carmustine | 3.3 mg/mL |
| Cyclophosphamide | 20 mg/mL |
| Doxorubicin hydrochloride | 2 mg/mL |
| Etoposide | 20 mg/mL |
| Fluorouracil | 50 mg/mL |
| Paclitaxel | 6 mg/mL |
| Thiotepa | 10 mg/mL |
| Two additional drugs | Selected and concentration documented by the manufacturer |
A glove can exceed 240 minutes for several drugs yet show a much shorter time for carmustine or thiotepa. FDA 510(k) summaries provide real examples of this pattern. That does not make one drug result an error; it demonstrates why the report must be read drug by drug.
Why carmustine and thiotepa often stand out
Permeation depends on the interaction between the drug formulation and the glove polymer. Molecular size, polarity, solvent system, concentration and affinity for the polymer all influence absorption and diffusion.
Carmustine and thiotepa frequently produce shorter breakthrough times in published glove data. A report that lists only the best-performing drugs can therefore create a misleading impression.
The practical rule is:
> Never transfer a breakthrough time from one drug to another, even when both are tested on the same glove.
The same rule applies to concentration. A test performed at one documented concentration does not automatically establish performance for a different formulation or mixture.
ASTM D6978 versus ASTM F739
ASTM F739 is a general method for measuring permeation of liquids and gases through protective-clothing materials. ASTM D6978 is tailored to medical gloves and chemotherapy drugs.
| Comparison point | ASTM D6978 | ASTM F739 |
|---|---|---|
| Primary use | Chemotherapy-drug permeation through medical gloves | General chemical permeation through protective materials |
| Test temperature | 35 +/- 2 degrees C | Conditions selected under the method and report |
| Breakthrough-rate threshold | 10 ng/cm2/min | Common standardized threshold of 100 ng/cm2/min under current general reporting practice |
| Specimen | Thinnest palm or cuff area | Representative protective-material specimen |
| Reporting emphasis | Minimum breakthrough time for each drug | Breakthrough and permeation parameters for the chemical/material pair |
The D6978 threshold is ten times lower than 100 ng/cm2/min. This more sensitive endpoint reflects the hazardous nature of antineoplastic drugs. A general chemical-resistance result should not be substituted for drug-specific D6978 data when a chemotherapy claim is required.
How to read a test report correctly
A useful ASTM D6978 report should identify:
- the exact glove model, material, size, color and lot;
- measured thickness at the tested specimen location;
- the standard edition;
- every drug name and its test concentration;
- the number of specimens or replicates;
- individual or minimum breakthrough times;
- results reported as minutes or greater than 240 minutes;
- the analytical method and detection capability; and
- the laboratory and test date.
Minimum values matter more than a convenient average. If three specimens break through at 38, 52 and 61 minutes, the test report should not allow a reader to mistake the 50-minute average for the earliest observed protection limit.
The report should also connect the tested samples to production. A test on a thicker development sample does not automatically support a thinner commercial glove, even if both are described as nitrile.
Test time is not the same as safe wear time
Laboratory breakthrough time is one input to a risk assessment, not a ready-made glove-change schedule. Real use adds factors that a flat, stationary permeation cell does not reproduce:
| Workplace factor | Possible effect |
|---|---|
| Stretch over fingers and knuckles | Locally reduces film thickness |
| Flexing and gripping | Applies repeated mechanical stress |
| Abrasion or small punctures | Creates new pathways through the film |
| Contact with alcohol or cleaning products | May alter the polymer before drug exposure |
| Drug mixtures or excipients | Can behave differently from the tested solution |
| Heat and perspiration | Change the conditions at the hand-glove interface |
Policies for handling hazardous drugs may also require double gloving and scheduled changes. Those decisions should follow the healthcare facility’s risk assessment, product instructions and applicable regulations or professional guidance, not a blog or a single breakthrough number.
What “tested for use with chemotherapy drugs” should mean
The phrase should be supported by product-specific ASTM D6978 results. It should not mean merely that the glove is nitrile, thick, powder-free or resistant to an unrelated solvent.
A transparent claim names the drugs tested and provides their minimum breakthrough times. It also highlights short results rather than hiding them behind a broad statement.
The central lesson is straightforward: ASTM D6978 measures drug-specific molecular permeation under controlled continuous contact. The earliest result for the relevant drug matters, and a value greater than 240 minutes is only a statement about the four-hour test window. Proper glove selection combines that evidence with task conditions, inspection, changing practices and the complete hazardous-drug program.
Technical references
- CDC PPE-Info: ASTM D6978-05(2023) standard details
- FDA recognized consensus standards listing ASTM D6978-05(2023)
- FDA product classification for medical gloves with chemotherapy labeling claims
- NIOSH discussion of chemotherapy-glove permeation standards
- FDA 510(k) example with drug-specific minimum breakthrough times
This article provides general technical education. It does not certify any ArmorHive glove for chemotherapy use and does not replace the complete standard, product-specific test report or an occupational risk assessment.